Implantable cardiac pacemaker electromagnetic compatibility testing in a novel security system simulator.

Wolfgang Kainz1, Jon P Casamento, Paul S Ruggera

  • 1Food and Drug Administration, Center for Devices and Radiological Health, Rockville, MD 20851, USA. wxk@cdrh.fda.gov

IEEE Transactions on Bio-Medical Engineering
|March 12, 2005
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Computational study of the effects of orthopedic plates on gradient-induced peripheral nerve stimulation under MRI using electromagnetic and neurophysiological modeling.

Magnetic resonance in medicine·2025
Same author

RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T.

Magnetic resonance in medicine·2024
Same author

A locally modified phantom model for assessing the worst-case heating configuration of orthopedic implants under MRI.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference·2023
Same author

RF-induced heating for active implantable medical devices in dual parallel leads configurations at 1.5 T MRI.

Magnetic resonance in medicine·2023
Same author

Infrared Thermography for Measuring Elevated Body Temperature: Clinical Accuracy, Calibration, and Evaluation.

Sensors (Basel, Switzerland)·2022
Same author

MR Conditionality of Abandoned Leads from Active Implantable Medical Devices at 1.5T.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference·2021

A new simulator accurately tests active implantable medical devices (AIMDs) against security systems, addressing over 100 past interference incidents. Pulsed magnetic fields pose a greater risk to pacemakers than sinusoidal fields.

Area of Science:

  • Biomedical Engineering
  • Electromagnetic Compatibility (EMC)
  • Medical Device Safety

Background:

  • Over 100 incidents reported over 17 years show interference between active implantable medical devices (AIMDs) and security systems.
  • A lack of standardized testing methods hinders the evaluation of AIMD electromagnetic compatibility (EMC) with security systems.

Purpose of the Study:

  • To develop and evaluate a novel simulator for EMC testing of AIMDs against electromagnetic fields from security systems.
  • To provide a standardized method for assessing the interference potential between AIMDs and walk-through metal detectors (WTMDs).

Main Methods:

  • A security system simulator was developed and evaluated for field homogeneity, signal distortion, and magnetic field strength.
  • Small three-axis probes and a scanning system characterized fields from 12 WTMD types.

Related Experiment Videos

  • Tests were conducted on four pacemakers using a saline phantom, correlated with "in air" tests.
  • Main Results:

    • The simulator accurately mimicked pacemaker interference observed in real WTMDs.
    • Interference thresholds indicated pulsed magnetic fields are more likely to interfere with pacemakers than sinusoidal fields.
    • Evaluated simulator performance regarding field homogeneity, signal distortion, and maximum magnetic field strength.

    Conclusions:

    • The developed security system simulator is a valuable tool for evaluating AIMD compatibility with security systems.
    • Early identification of incompatible WTMD and AIMD combinations is facilitated by this simulator.
    • The findings highlight the increased risk of pulsed magnetic fields causing pacemaker interference.