[Outlier diagnosis and calibration model optimization for near infrared spectroscopy analysis]
Shun-geng Min1, Ning Li, Ming-xiang Zhang
1College of Science, China Agricultural University, Beijing 100094, China.
Abstract:
Outlier diagnosis is a very important step in building near infrared calibration model. Data outlier includes spectral outlier and chemical value outlier. Mahalanobis' distance, ratio of spectral residual and spectral variable leverage test were used to evaluate sample spectral outlier. Cook's distance and the ratio of sample square error of chemical value and predict value to the mean square error of calibration set were used to test chemical value outlier. Three calibration models of protein content of 50 wheat samples, protein content of 90 corn samples and cyclohexane content of four compounds mixture were investigated. It is demonstrated that outlier test is very helpful for optimizing near infrared calibration model.
More Related Videos
06:50O-cresol Concentration Online Measurement Based On Near Infrared Spectroscopy Via Partial Least Square Regression
Published on: November 8, 2019
07:11ARL Spectral Fitting as an Application to Augment Spectral Data via Franck-Condon Lineshape Analysis and Color Analysis
Published on: August 19, 2021
Related Concept Videos
Outliers and Influential Points
Detection of Gross Error: The Q Test
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Calibration Curves: Correlation Coefficient
Quantifying and Rejecting Outliers: The Grubbs Test
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...