Related Experiment Video
Updated: Aug 19, 2026

Tools for Surface Treatment of Silicon Planar Intracortical Microelectrodes
Published on: June 8, 2022
Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coating
Lucien Reclaru1, Pierre-Yves Eschler, Reto Lerf
1PX Holding S.A., Bldv des Eplature 42, 2304 La Chaux-de-Fonds, Switzerland. lucien.reclaru@pxgroup.com
Abstract:
The corrosion behavior of CoCrMo implants with rough titanium coatings, applied by different suppliers by either sintering or vacuum plasma spraying, has been evaluated and compared with uncoated material. The open-circuit potential, corrosion current and polarization resistance were determined by electrochemical techniques. The Co, Cr and Ti ions released from the samples into the electrolyte during a potentiostatic extraction technique were analyzed using ICP-MS. The Ti coatings from the different suppliers showed a different porous morphology, and the implants exhibited a distinct corrosion activity, underlining the importance of the coating process parameters. Among the titanium coated samples, the one with the sintered overcoat turned out to be the most resistant. Yet, on an absolute scale, they all showed a corrosion resistance inferior to that of uncoated CoCrMo or wrought titanium.
More Related Videos
Related Concept Videos
Microbial Corrosion
Corrosion
Electrodeposition
Electrodeposition can...
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...