Validation of timed micturition--a comparison of flow rates measured at home manually and electronically
Bengt Folkestad1, Anders Spångberg
1Department of Urology, Faculty of Health Sciences, Linköping University, Linköping, Sweden. bengt.folkestad@lio.se
Objective:
To compare and evaluate manually performed timed micturition in a domestic setting with registrations made using a home flowmeter.
Material And Methods:
Forty-six randomly selected men without voiding symptoms performed timed micturition at home both by hand and with a Da Capo home flowmeter. Statistical comparison using the Wilcoxon signed-rank test and regression analysis of the differences was performed to evaluate whether the two methods were comparable.
Results:
In two cases, there was a large discrepancy between the two methods. For the remaining 44 men, there were no statistically significant differences between the two methods. The median (range) timed micturition was 7 s (3-35 s) for manual measurements and 6.5 s (3.5-28 s) for electronic measurements. For each individual, the difference between the manually and electronically obtained medians was calculated. The average individual difference was 0.06 s (SD 1.6 s).
Conclusion:
Approximately 95% of men are able to measure timed micturition at home manually without statistically or clinically significant differences from electronically obtained measurements. One has to be aware that in a small proportion of cases timed micturition measured at home may be falsely long.
More Related Videos
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Pipe Flowrate Measurement
The orifice meter is a simple,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
The Micturition Reflex
The process begins with bladder filling, where the bladder wall stretches as urine accumulates. This stretching activates the urine storage reflex, mediated by the sacral spinal segments and the pontine storage center. Efferent sympathetic impulses stimulate the detrusor muscle to relax and the internal urethral sphincter to contract, facilitating urine...
Pipe Flowrate Measurement: Problem Solving