The performance of air-turbine handpieces in general dental practice
D M Monagahn1, N H F Wilson, B W Darvell
1The Buxton Dental Practice, Buxton, Derbyshire, UK. dmm121@hotmail.com
Objective:
To investigate variation in performance measures of fibre-optic, high-speed air-turbine handpieces during the course of daily use in general dental practice.
Materials And Methods:
Four groups of five new high-speed fibre-optic handpieces were used in the routine treatment of patients over a period of 30 months by four general dental practitioners in two dental practices: Groups A, B: Super-Torque Lux 3 650B (KaVo, Biberach, Germany); Group C: BORA 898LE (BienAir SA, Bienne, Switzerland) and Group D: Toplight (W&H Dentalwerk, Burmoos, Austria). The dental practice teams had been rehearsed in the procedures to be followed before starting the study. Each dentist used the handpieces in strict rotation, while the groups were rotated monthly between practitioners. Four performance characteristics were measured before use, then at regular intervals: free-running speed (Hz) and bearing resistance (microNm) were measured using a purpose-built testing machine (Darvell-Dyson); illuminance (lux) and sound pressure level (dB(A)) were also measured. Handpieces were cleaned and lubricated in accordance with manufacturers' directions; all were autoclaved wet at 134 degrees C for three minutes.
Results:
Free-running speed showed an initial increase after use for Groups A, B and C, which may be associated with a decrease in bearing resistance. All handpieces in Group C suffered bearing failure between months 21 and 23, preceded by a substantial increase in noise, while those in Group D suffered failure of the fibreoptic system between months 18 and 24. Other deterioration due to use was identified but Groups A, B and D were still in use at month 30.
Conclusions:
Variation in free-running speed, bearing resistance, illuminance and sound pressure level can be used effectively to monitor changes in air-turbine handpieces due to normal use. Although an increase in bearing resistance is associated with decreasing free-running speed, noise appears to be a useful indicator of imminent bearing failure. Assiduous adherence to manufacturers' directions for cleaning and lubrication may have contributed to increased bearing life.
Related Concept Videos
Tracheostomy Suctioning II: Procedure
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Suctioning the Nasopharyngeal Airway
Equipment Required
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...