Long-term Blood Pressure Measurement in Freely Moving Mice Using Telemetry
Published on: May 17, 2016
Repeatability of automated ambulatory blood pressure measurements
L M Prisant1, A A Carr, P B Bottini
1Section of Hypertension, Medical College of Georgia, Augusta 30912-3150.
Background:
The repeatability of 24-hour automatic ambulatory blood pressure measurements recorded by noninvasive equipment (Del Mar Pressurometer IV) was assessed to determine the intrapatient variability of this test.
Methods:
The usual antihypertensive medications of 73 patients with documented essential hypertension (supine diastolic blood pressure of 95 to 119 mm Hg) were withdrawn, and the patients were treated with placebo medication for 6 weeks. At the end of the placebo period, ambulatory blood pressure measurements of each patient were recorded every 15 minutes for 24 hours on two separate occasions 1 week apart.
Results:
There was no significant difference in either the 24-hour systolic or diastolic blood pressure for the entire group between weeks. A mean difference for individual patients between the first and second recording within 5 mm Hg was observed in 49.3% and 52.1% of patients for 24-hour systolic and diastolic blood pressure, respectively. The correlation coefficient for 24-hour systolic blood pressure was greater than 24-hour diastolic blood pressure (r = .87 vs r = .67). A difference greater than 18.1/14.9 mm Hg for systolic/diastolic blood pressure would be required to assign a significant (P less than .05) change in blood pressure between two recordings in the same patient.
Conclusions:
These data quantify the usefulness of 24-hour ambulatory blood pressure measurements for a group of subjects. However, mean 24-hour ambulatory blood pressure varies significantly for individuals. Intrapatient variability may limit the usefulness of a single 24-hour ambulatory recording for an individual patient and suggests the need for more than one measurement to establish a level of blood pressure.
More Related Videos
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.