Related Experiment Video
Updated: Aug 18, 2026

Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
Decreased intermediate term patency of automated proximal anastomoses evaluated by sequential ultrafast CT
Markus Dietrich1, Sven Martens, Melanie Kohlert
1Department of Thoracic and Cardiovascular Surgery, JW Goethe University Clinic, Frankfurt, Germany. md.md@gmx.de
Objective:
The Symmetrytrade mark aortic connector creates proximal anastomoses of saphenous vein grafts using a nitinol implant. The device avoids partial clamping and thus possibly reduces neurologic complications. To evaluate graft patency, a single surgeon randomised study was performed in our institution.
Methods:
Seventy-seven patients were randomised either to automated proximal anastomoses (group I, n=39, 61 vein grafts connected using the aortic connector, 47 as single, 15 as sequential bypasses) or controls (group II, n=38, 62 proximal anastomoses handsewn, 46 as single, 16 as sequential bypasses). Ultrafast CT-scans were performed on postoperative day 5 in 34 patients of group I and 16 patients of group II to evaluate early graft patency. Intermediate term patency was evaluated with ultrafast CT-scan in 30 patients of group I (46 grafts) and 25 patients of group II (39 grafts) 1 year after the operation.
Results:
Two early graft occlusions were detected in group I (3.8%). In group II all evaluated grafts were patent 5 days after surgery. 11.4 months after surgery, seven out of 46 grafts were found occluded (15.2%) in group I. In the control group, only one occlusion was detected (2.6%).
Conclusions:
We observed a trend towards an increased occlusion rate 1 year after surgery with automated connector devices. For evaluation of long-term patency larger patient groups have to be evaluated. Other benefits of these devices have to be proven to promote their clinical application.