Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Polyp detection with MDCT: a phantom-based evaluation of the impact of dose and spatial resolution
Abdülvahit Ozgün1, Erik Rollvén, Lennart Blomqvist
1Department of Medical Physics, S-171 76 Karolinska University Hospital, Stockholm, Sweden. vahit123@hotmail.com
Objective:
The objective of our study was to evaluate the impact of dose and spatial resolution on the detection of colonic polyps using a 4-MDCT scanner.
Materials And Methods:
Twenty-four latex phantoms that simulate the large bowel and contain artificial polyps of different sizes and shapes were constructed. The polyps were divided into three size groups (diameter, 0-2, 2-5, and 5-10 mm) and were classified into four shape groups: pedunculated; broad-based; ulcerated or depressed; and sessile or flat. The colon phantoms were submerged in a water tank and scanned on a 4-MDCT scanner using 12 protocols with various settings of slice thickness, pitch, and tube current. The images were independently evaluated by three radiologists using axial 2D multiplanar reconstruction images and a 3D surface-rendering technique (fly-through).
Results:
At a constant dose (i.e., dose-length product [DLP]), the polyp detection rate increased with increasing axial spatial resolution. For the standard protocol (2.50-mm slice thickness, 1.5 pitch), the detection rate for all polyp sizes decreased from approximately 70% at 100 mA to 55% at 40 mA. Between a 60- and 100-mA tube current, the detection rate for the largest polyps (> 5 mm) was almost constant, close to 90%.
Conclusion:
The detection of polyps in the large bowel using a standard protocol can be improved without dose penalty by increasing the axial spatial resolution of the image acquisition and adjusting the tube current setting. If the analysis can be restricted to polyps larger than 5 mm, the dose can be substantially reduced without compromising the detection rate.