Related Experiment Video
Updated: Aug 18, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Determination of oxycodone and hydrocotarnine in cancer patient serum by high-performance liquid chromatography with
Hideya Kokubun1, Makiko Ouki, Motohiro Matoba
1Department of Pharmacy, Kitasato University Hospital, 1-15-1, Kitasato, Sagamihara, Kanagawa 228-8555, Japan. kokubun@nc.kitasato-u.ac.jp
Abstract:
We developed an HPLC procedure using electrochemical detection for the quantitation of oxycodone and hydrocotarnine in cancer patients serum. An eluent of methanol:acetonitrile:5 mM pH 8 phosphate buffer (2:1:7) was used for the mobile phase. The calibration curve was linear in the range from 10 ng/mL to 100 ng/mL. The recovery of oxycodone and hydrocotarnine was 97.2% and 90.5%, respectively. The relative standard deviations within-runs and between-runs for the assay of oxycodone or hydrocotarnine were less than 4.8%. The method developed here was better than the method reported previously.
More Related Videos
08:57Sample Extraction and Simultaneous Chromatographic Quantitation of Doxorubicin and Mitomycin C Following Drug Combination Delivery in Nanoparticles to Tumor-bearing Mice
Published on: October 5, 2017
10:38Simultaneous Quantification of Selected Kynurenines Analyzed by Liquid Chromatography-Mass Spectrometry in Medium Collected from Cancer Cell Cultures
Published on: May 9, 2020
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: