Related Experiment Video
Updated: Aug 18, 2026

Extraction of Non-Protein Amino Acids from Cyanobacteria for Liquid Chromatography-Tandem Mass Spectrometry Analysis
Published on: December 9, 2022
Dynamic ultrasound-assisted extraction of colistin from feeds with on-line pre-column derivatization and liquid
S Morales-Muñoz1, M D Luque de Castro
1Department of Analytical Chemistry, University of Córdoba, Marie Curie Building Annex, Campus of Rabanales, E-14071 Córdoba, Spain. a62momus@uco.es
Abstract:
A dynamic ultrasound-assisted extraction (UAE) method with on-line pre-column derivatization/high performance liquid chromatography (HPLC) and fluorimetric detection is proposed for the analysis of colistin in feed. A flow injection manifold is used for the development of the extraction and derivatization steps and for interfacing them with the separation/detection step, thus providing an on-line approach with the advantage of minimum sample handling. The derivatization was performed with ortho-phthaldialdehyde and 2-mercaptoethanol. The optimum conditions for colistin extraction and formation of the fluorescent derivative have been obtained by experimental design methodology. The use of a high-intensity probe sonication makes UAE an expeditious (7 min versus > 1 h) and efficient (93.1-98.2% versus 87.5-94% of recovery) alternative as compared with extraction using an ultrasonic bath. The within-laboratory reproducibility and repeatability, expressed as percentage of relative standard deviation, were 5.2 and 5.8, respectively.