Compliance and resistance levels and unloading in proportional assist ventilation
Jaana A Leipälä1, Shiho Iwasaki, Silke Lee
1Children Nationwide Regional Neonatal Intensive Care Centre, King's College Hospital, London SE5 9RS, UK.
Abstract:
During proportional assist ventilation (PAV), resistive and elastic unloading relieve the work of breathing. Excessive unloading, however, results in resonant oscillations and runaway pressures. Our aim was to determine the appropriate levels of unloading that could be applied to clinical practice. A lung model, resistance (50 or 150 cmH(2)O l(-1) s(-1)) and compliance (0.4 or 0.8 ml/cmH(2)O), was used. The volumes and airway pressures delivered by the ventilator at varying levels of resistive and elastic unloading and simulated breaths were recorded. Oscillations in airway pressure only occurred when the level of resistive unloading exceeded the model's resistance. When the level of unloading fully compensated for the model's compliance, peak inflating pressures greater than 40 cmH(2)O were delivered; peak pressure limits of 20 cmH(2)O, however, resulted in very short (0.2 s or less) inflation times. High peak pressures were not delivered if the level of elastic unloading used was limited to that which reduced the model's elastance to that of a 'normal lung'. In conclusion, these results suggest that when using PAV, it is important to assess the compliance and resistance of the infant and endotracheal tube, so that levels of unloading that fully compensate for the resistance and compliance levels can be avoided.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)