Reassessing the clinical efficacy of two MR quantitative DSC PWI CBF algorithms following cross-calibration with PET

Jean J Chen1, Richard Frayne, Michael R Smith

  • 1Department of Electrical and Computer Engineering, University of Calgary, 2500 University Drive, NW, Calgary, Alberta T2N1N4, Canada.

Insights

Clinical cerebral blood flow (CBF) maps from DSC MR perfusion imaging require rescaling for accurate interpretation. This process impacts how deconvolution algorithm performance is assessed, emphasizing consistency over absolute accuracy with varying mean transit times (MTT).

Area of Science:

  • Neuroimaging
  • Medical Physics
  • Radiology

Background:

  • Clinical cerebral blood flow (CBF) maps are derived from dynamic-susceptibility contrast (DSC) magnetic resonance (MR) perfusion imaging.
  • Current cross-calibration with PET studies involves rescaling MR CBF values to a standard normal white matter value (22 ml/100 g/min).

Purpose of the Study:

  • To evaluate the impact of CBF map rescaling on clinical interpretation.
  • To assess how rescaling affects the evaluation of deconvolution algorithm performance.

Main Methods:

  • Analysis of singular-value decomposition (SVD) and frequency-domain modeling algorithms for CBF estimation.
  • Examination of the effects of rescaling on CBF estimates across varying mean transit times (MTT).

Main Results:

  • Rescaling can lead to over-estimated CBF maps for MTT >4.8 s when using SVD algorithms.
  • Both SVD and frequency-domain algorithms show reduced sensitivity to recirculation artifacts after rescaling.
  • The relative accuracy of CBF estimates becomes more critical than inherent accuracy after rescaling.

Conclusions:

  • Rescaling procedures significantly alter CBF map interpretation and algorithm assessment.
  • Developing deconvolution algorithms with consistent accuracy across different MTTs is crucial when rescaling is applied.