Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
New flexible polymeric heart valve prostheses for the mitral and aortic positions
Sabine H Daebritz1, Bernd Fausten, Benita Hermanns
1Department of Cardiac Surgery, Ludwig-Maximilians-University, Munich, Germany. sabine.daebritz@med.uni-muenchen.de
Objective:
Current prosthetic heart valves necessitate permanent anticoagulation or have limited durability and impaired hemodynamic performance compared with natural valves. We report in vivo and in vitro results with new polymeric valve prostheses that have a special design for the mitral and aortic positions. The aims are improved durability and elimination of the need for permanent anticoagulation.
Methods:
The mitral and aortic prostheses (Adiam Life Science, Erkelenz, Germany) are made entirely of polycarbonate urethane (PCU). The bileaflet asymmetric mitral valve mimics natural, nonaxial inflow, which creates a left ventricular vortex, saving energy for systolic ejection of blood. The trileaflet aortic prosthesis has diminished pressure loss and reduced stress and strain peaks at the commissures. The valves were subjected to long-term in vitro testing and in vivo testing in a growing calf model (20 weeks; 7 mitral and 7 aortic valves) with comparison with 2 commercial bioprostheses (7 mitral, 2 aortic). Two-dimensional echocardiography was performed after implantation and prior to sacrifice with autopsy and valve examination.
Results:
In vitro durability of the PCU valves was proved up to 20 years. In vivo durability and hemodynamics were superior to those of all bioprostheses. Survival of PCU valves versus bioprostheses was 7 versus 2 mitral valves and 5 versus 0 aortic valves, respectively. Two animals with PCU aortic valves died of pannus overgrowth that caused severe left ventricular outflow tract obstruction without changes in the valves. Degeneration and calcification were mild (mitral) and moderate (aortic) in PCU valves but were severe in biological valves. There was no increased thrombogenicity of the PCU valves compared with bioprostheses.
Conclusion:
The new flexible polymeric aortic and mitral valve prostheses were superior to current bioprostheses in animal testing.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
06:56An Ex Vivo Porcine Model for Hydrodynamic Testing of Experimental Aortic Valve Procedures and Novel Medical Devices
Published on: August 25, 2023
Related Concept Videos
Mitral Stenosis III: Medical Management
Mitral Valve Prolapse I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Regurgitation III: Medical Management
Mitral Valve Prolapse II: Assessment and Management
Mitral Stenosis I: Introduction