Inter-instrumental variation of skin capacitance measured with the Corneometer
Ken-Ichiro O'goshi1, Jørgen Serup
1Department of Dermatology, Bispebjerg Hospital, Copenhagen, Denmark. kenmockjp@yahoo.co.jp
Background/Purpose:
Measurement of skin surface and stratum corneum (SC) hydration during clinical and/or experimental trials needs devices with acceptable reproducibility and sensitivity under conditions ranging from increased and normal to low hydration. A previous study comparing Corneometer instruments (European group for Efficacy Measurement of Cosmetics and Other products--EEMCO) used for measurement of electrical capacitance of skin indicated a major difference among Corneometer instruments. The aim of this study was to assess three inter-instrumental similar Corneometer instruments (two pieces of CM820 and one CM810) in normal skin. We named them CM-A(CM820), CM-B(CM820), and CM-C(CM810).
Methods:
The hydration state of SC measured as electrical capacitance of six body sites were measured in 53 subjects with three Corneometer instruments.
Result:
We found that the Corneometer instruments displayed close capacitance levels. When one Corneometer was plotted against another the regression line indicated a good correlation among instruments albeit a major and random disagreement could appear in individual sites as an exception.
Conclusion:
Three Corneometer instruments evaluated in this study gave close measurements and correlated well. Nevertheless, pretest validation of instruments should be undertaken in multicenter studies where capacitance data are compared or pooled, and concordance among instruments should be documented prior to study.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equivalent Capacitance
The following strategies are adopted to calculate...
Sensory Functions of the Skin
There are two main categories of receptors on the skin: capsulated and non-capsulated. The non-capsulated ones are mainly the pain receptors. The capsulated ones can be further categorized based on the...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Equivalent Capacitance