Related Experiment Video
Updated: Aug 18, 2026

Quantitative Measure of Lung Structure and Function Obtained from Hyperpolarized Xenon Spectroscopy
Published on: November 10, 2023
Measurements and modeling of long range 3He diffusion in the lung using a "slice-washout" method
Stanislao Fichele1, Martyn N J Paley, Neil Woodhouse
1Academic Unit of Radiology, The University of Sheffield, Royal Hallamshire Hospital, Glossop Road, Sheffield S10 2JF, UK.
Abstract:
In healthy lung tissue, pulsed-gradient-spin-echo (PGSE) methods reveal apparent diffusion coefficients (ADC) of the order 0.20 cm2 s(-1); for diffusion times of approximately 2 ms. For these short diffusion times the ADC is only sensitive to structures approximately (2Dt)1/2 approximately 0.6mm in size. Recent work, using magnetic tagging of the longitudinal magnetization has revealed much smaller ADC values for longer length scales. In this work, the in vivo ADC from within the air-spaces, was measured using a new technique. The signal from a series of images was analyzed from a slice that was repeatedly imaged. Diffusion tends to "top-up" the non-renewable polarization within the slice, which leads to a non-exponential decay in image signal. Image data were compared to 1D finite-difference simulations of diffusion to calculate a long range ADC value. The results yield values of the order 0.034 cm2 s(-1), which are nearly an order of magnitude smaller than those reported by PGSE measurements at shorter diffusion times.
