A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Intra- and interobserver variability during in vitro validation of two novel methods for intra-abdominal pressure
Tom J R De Potter1, Hilde Dits, Manu L N G Malbrain
1Intensive Care Unit, Campus Stuivenberg, ZiekenhuisNetwerk Antwerpen, Lange Beeldekensstraat 267, 2060 Antwerpen 6, Belgium.
Objective:
Intra-abdominal pressure (IAP) measurement techniques vary in automaticity and reproducibility. This study evaluated the intra- and interobserver variability of two new IAP measurement techniques.
Methods:
A one-half open, 30-l container was used with two IAP catheters (Foley Manometer, balloon-tipped catheter) contained in a 100-ml infusion bag. To simulate intra-abdominal hypertension the container was filled with water using 5-cm increments (0-25 cmH2O). Pressure was determined by observers using the Foley Manometer and simultaneously recorded using an IAP monitor. Observers were blinded to the reference levels.
Results:
Fifteen observers conducted three pressure readings at each of the six pressure levels with the Foley technique, giving 270 readings. These were paired with the automated monitor readings and the height of the water column. The intra- and interobserver coefficients of variation were low for both methods. The Spearman correlation coefficient was higher than 0.9 for all paired measurements and Bland-Altman analysis comparing the reference H2O column to both measurement techniques showed a very good agreement at all pressure intervals (bias 0.1+/-0.6 cmH2O) and a consistent, low underestimation of the reference water column pressure by both techniques.
Conclusions:
Both the Foley Manometer and the IAP monitor are reliable and reproducible methods to measure IAP in this in vitro model. The coefficient of variation for each technique is low and decreases with increasing IAP, the monitor giving more reproducible results than the Foley Manometer.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...