Related Experiment Video
Updated: Aug 18, 2026

Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
Published on: December 22, 2023
[Determination of six trace elements in steels by ICP-AES with the changed internal standard]
Abstract:
This paper studies the direct determination of P, Si, Mn, Cu, Cr and Ni in steels with the changed internal standard. The standard curves are standardized with FULL STANDARDIZATION, and the results of the sample are calibrated with TYPE STANDARD in time. The signal intensity influenced by RF power, observation height and sample gas folw is discussed. Results obtained for the determination of every element in steels are satisfactory. The RSD is less than 0.5% and the recovery is about 96-103%. The detection limit is about 0. 002-0.07%.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Overview