Colorimetric carbon dioxide detector to determine accidental tracheal feeding tube placement
Daniel W Howes1, Eric S Shelley, William Pickett
1Empire 3, Department of Emergency Medicine, Queen's University, Kingston, Ontario, Canada. danielwilliamhowes@hotmail.com
Purpose:
To determine the accuracy of colorimetric CO2 detection compared to the reference standard two-step radiological confirmation of feeding tube position.
Methods:
A prospective study was conducted with patients presenting to a 21-bed medical-surgical intensive care unit. An adapter was developed using an endotracheal tube adapter to connect a colorimetric CO2 detector to a feeding tube in an airtight manner. In part I of the study a feeding tube connected to the colorimeter was inserted into the endotracheal tubes of ten ventilated patients to test the device's ability to detect tracheal placement. In part II patients undergoing feeding tube insertion had tube position confirmed with the colorimeter as well as the reference standard two-step x-ray.
Results:
In phase I the colorimeter correctly identified tracheal placement in all ten patients. In phase II 93/100 procedures ultimately were eligible; the colorimeter had a sensitivity of 0.88 (95% confidence interval: 0.65-1.00) and specificity of 0.99 (0.97-1.00). The device missed one of the eight tracheal placements. Agreement between the colorimeter and two-step x-ray interpretations was excellent (Kappa 0.86; standard error 0.10).
Conclusion:
We describe a novel, convenient method to confirm esophageal feeding tube placement. The device is easily assembled and inexpensive, but should not be reused. Colorimetric determination of tracheal feeding tube placement with this device has excellent agreement with the reference standard two-step radiological technique.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Gas Chromatography: Types of Detectors-II
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...