Does thoracic bioimpedance accurately determine cardiac output in COPD patients during maximal or intermittent
Valerie Bougault1, Evelyne Lonsdorfer-Wolf, Anne Charloux
1Service des Explorations Fonctionnelles Respiratoires et de l'Exercice, Hôpital Civil, BP 426, 67091 Strasbourg Cedex, France.
Study Objectives:
The monitoring of cardiac output (CO) during exercise rehabilitation in patients with COPD, often including strenuous exercise, is advisable. Invasive methods (thermodilution, Fick method) are accurate, but for clinical routine use noninvasive CO estimation is required. We have shown that impedance cardiography (Physio Flow; Manatec Biomedical; Macheren, France) is reliable in COPD patients at rest and during a recumbent, light-intensity exercise. The aim of our study was to evaluate the validity of this noninvasive device in COPD patients during a maximal incremental exercise test (IET) and also during a strenuous intermittent work exercise test (IWET).
Design:
Prospective comparative study of the impedance cardiograph vs the direct Fick method applied to oxygen.
Patients:
Eight patients with moderate-to-severe COPD (59 +/- 6 years old; FEV(1), 38 +/- 15% predicted; residual volume, 194 +/- 64% predicted) [mean +/- SD].
Measurements And Main Results:
Forty-nine simultaneous measurements of CO by means of the direct Fick method (COfick) and CO measured by the impedance cardiograph (COpf) were obtained during the IET, and 108 measurements were made during the IWET. The correlation coefficients between the two measurements were r = 0.85 and r = 0.71 for the IET and the IWET, respectively. COpf was higher than COfick. The difference between the two methods was 3.2 +/- 2.9 L/min during the IET and 2.5 +/- 2.1 L/min during the IWET. Expressed as a percentage of the mean of the two measurements, this corresponded to 31 +/- 21% and 25 +/- 20%, respectively.
Conclusions:
The relatively high number of values differing by > 20% precludes the use of impedance cardiography in clinical routine in such a difficult setting (hyperinflated patients and intense exercise).
More Related Videos
08:44Dual Test Gas Pulmonary Diffusing Capacity Measurement During Exercise in Humans Using the Single-Breath Method
Published on: February 2, 2024
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...
Exercise and Cardiovascular Response
Light to moderate physical activity initiates a series of interconnected responses in the body. The heart rate modestly increases in anticipation of the workout, followed by widespread vasodilation as oxygen consumption by skeletal muscles increases. This results in decreased peripheral resistance, increased capillary blood flow, and accelerated...
Chronic Obstructive Pulmonary Disease-I: Introduction
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...