Validation of direct intraabdominal pressure measurement using a continuous indwelling compartment pressure monitor
A J Brooks1, A Simpson, M Delbridge
1Department of Surgery, Queens Medical Centre, Nottingham, United Kingdom.
Background:
According to recommendations, intraabdominal pressure should be monitored every 8 hours for patients at high risk of abdominal compartment syndrome. Continuous intraabdominal pressure monitoring may be valuable for these patients.
Methods:
For 15 patients undergoing laparoscopic surgery, a pressure monitor was introduced after formation of pneumoperitoneum. During the procedure, the laparoscopic insufflator pressure was varied. The pressure monitor values and the time to equilibrium were recorded.
Results:
Altogether, 152 pressure recordings were taken for the patients studied. The measurements from the insufflator and pressure monitor were compared using a Bland-Altman plot. The mean difference between the techniques was 0.04 +/- 3.8, and 95% of the points from the pressure monitor were within two standard deviations of the mean difference. Pressure changes were essentially "real time."
Conclusions:
The intracompartmental pressure monitor provides accurate, rapid, and direct measurement of intraabdominal pressure, and may be a useful tool for continuous intraabdominal pressure measurement among patients at risk of abdominal compartment syndrome.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Increased Intracranial Pressure l: Introduction