Fire in the operating room during tracheotomy: a case report
Related Experiment Video
Updated: Aug 18, 2026

Endotracheal Intubation via Tracheotomy and Subsequent Thoracotomy in Rats for Non-Survival Applications
Published on: March 15, 2024
Dona H Paugh1, Kathryn W White
1Veterans Affairs Medical Center, Minneapolis, MN, USA.
Abstract:
We describe a patient who sustained facial burns during a tracheotomy. The electrosurgical unit indirectly started the fire during monitored anesthesia care when a high inspired oxygen concentration was being delivered to the patient by simple face mask. This case points out the need for prevention strategies, intraoperative vigilance, and quick intervention to prevent further patient injury any time the electrosurgical unit is used in an oxygen-enriched atmosphere.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...