Comparison of critical closing pressures extracted from carotid tonometry and finger plethysmography
Hung-Yi Hsu1, A-Ching Chao, Ying-Tsung Chen
1Department of Internal Medicine, Veterans General Hospital-Taichung, Taiwan.
Background:
The reliability of critical closing pressure (CrCP) estimates derived from peripheral blood pressure (BP) measurements is unclear. We attempted to evaluate the influences of peripheral circulation on determining CrCP.
Methods:
Twenty-five young healthy volunteers were studied. BP waves were obtained with plethysmography (Portapres) and carotid applanatory tonometry, respectively, for analysis. Transcranial Doppler was used to monitor cerebral flow velocity. Using linear regression analysis, beat-to-beat CrCP was calculated at rest, during voluntary hyperventilation and during 5% CO2 inhalation.
Results:
Twenty of 25 participants demonstrating satisfactory tonometric tracings for both tests were included in the analysis. The systolic BP measured using plethysmography was higher than that derived from tonometry (139.4 +/- 24.7 vs. 105.5 +/- 29.6, p < 0.001). CrCP values derived from tonometry were all positive and higher than CrCP values derived from plethysmography (62.9 +/- 19.9 vs. 11.1 +/- 17.8, p < 0.001). The changes in CrCP induced by 5% CO2 inhalation and hyperventilation had a correlation between two BP monitoring methods (r = 0.52, p = 0.001).
Conclusions:
Pressure waveform is an important determinant in calculating CrCP by linear regression analysis. The relative changes in CrCP induced by hemodynamic challenges remained a relevant indicator of cerebrovascular regulation regardless of the methods used for non-invasive BP recording.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: