Comparative scatter and dose performance of slot-scan and full-field digital chest radiography systems
Ehsan Samei1, Joseph Y Lo, Terry T Yoshizumi
1Duke Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, DUMC 3302, Durham, NC 27710, USA. samei@duke.edu
Purpose:
To evaluate the scatter, dose, and effective detective quantum efficiency (DQE) performance of a slot-scan digital chest radiography system compared with that of a full-field digital radiography system.
Materials And Methods:
Scatter fraction of a slot-scan system was measured for an anthropomorphic and a geometric phantom by using a posterior beam-stop technique at 117 and 140 kVp. Measurements were repeated with a full-field digital radiography system with and without a 13:1 antiscatter grid at 120 and 140 kVp. For both systems, the effective dose was measured on posteroanterior and lateral views for standard clinical techniques by using dosimeters embedded in a female phantom. The effective DQEs of the two systems were assessed by taking into account the scatter performance and the DQE of each system. The statistical significance of all the comparative differences was ascertained by means of t test analysis.
Results:
The slot-scan system and the full-field system with grid yielded scatter fractions of 0.13-0.14 and 0.42-0.48 in the lungs and 0.30-0.43 and 0.69-0.78 in the mediastinum, respectively. The sum of the effective doses for posteroanterior and lateral views for the slot-scan system (0.057 mSv +/- 0.003 [+/- standard deviation]) was 34% lower than that for the full-field system (0.086 mSv +/- 0.001, P < .05) at their respective clinical peak voltages (140 and 120 kVp, respectively). The effective DQE of the slot-scan system was equivalent to that of the full-field system in the lung region but was 37% higher in the dense regions (P < .05).
Conclusion:
The slot-scan design leads to marked scatter reduction compared with the more conventional full-field geometries with a grid. The improved scatter performance of a slot-scan geometry can effectively compensate for low DQE and lead to improved image quality.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...