Related Experiment Video
Updated: Aug 18, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Investigation of optimum energies for chest imaging using film-screen and computed radiography
I D Honey1, A Mackenzie, D S Evans
1Kings Centre for the Assessment of Radiological Equipment (KCARE), Kings College Hospital, London SE5 9RS, UK.
Abstract:
The purpose of the study was to compare the image quality of film-screen (FS) and computed radiography (CR) for adult chest examinations across a range of beam energies. A series of images of the CDRAD threshold contrast detail detection phantom were acquired for a range of tube potential and exposure levels with both CR and FS. The phantom was placed within 9 cm of Perspex to provide attenuation and realistic levels of scatter in the image. Hardcopy images of the phantom were scored from a masked light-box by two scorers. Threshold contrast indices were used to calculate a visibility index (VI). The relationships between dose and image quality for CR and for FS are fundamentally different. The improvements in VIs obtained using CR at 75 kVp and 90 kVp were found to be statistically significant compared with 125 kVp at matched effective dose levels. The relative performance of FS and CR varies as a function of energy owing to the different k-edges of each system. When changing from FS to CR, the use of lower tube potentials may allow image quality to be maintained whilst reducing effective dose. A tube voltage of 90 kVp is indicated by this work, but may require clinical verification.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...