Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart

Elizabeth B Fortescue1, Charles I Berul, Frank Cecchin

  • 1Arrhythmia Service, Department of Cardiology, Children's Hospital Boston and Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, USA.

Heart Rhythm
|April 27, 2005
PubMed

Insights

Pediatric pacemaker lead failures are common, particularly in young children with congenital heart defects and those with epicardial leads. Individualized care and considering transvenous systems can improve outcomes.

Area of Science:

  • Pediatric Cardiology
  • Biomedical Engineering
  • Medical Device Research

Background:

  • Pediatric pacemaker management is complex due to small patient size, congenital heart disease, and growth.
  • Identifying risk factors for pacemaker lead failure is crucial for improving outcomes in children.

Purpose of the Study:

  • To analyze pediatric pacemaker lead failure outcomes over a 22-year period.
  • To identify specific risk factors associated with lead failure in pediatric patients.

Main Methods:

  • Retrospective analysis of 1007 leads implanted in 497 pediatric patients (1980-2002).
  • Lead failures defined as repair, replacement, or abandonment due to mechanical or electrical issues.
  • Logistic regression and Cox analyses used to identify risk factors for lead failure.

Main Results:

  • 15% of leads (155) and 23% of patients (115) experienced lead failure.
  • Independent risk factors for lead failure included age <12 years, congenital heart defects, and epicardial lead placement.
  • Younger patients (<12) had more fractures; congenital heart defect patients had more exit block; epicardial leads had higher fracture/exit block rates.

Conclusions:

  • Pediatric pacing patients exhibit a high incidence of lead failures.
  • Lead failures are most prevalent in younger patients, those with congenital heart disease, and epicardial systems.
  • Individualized pacing strategies and increased use of transvenous systems (when feasible) are recommended to mitigate lead failure risks.
Abstract