Absent ventricular tachycardia detection in a biventricular implantable cardioverter-defibrillator due to intradevice
Joshua M Cooper1, William H Sauer, Ralph J Verdino
1University of Pennsylvania Health System, Philadelphia, 19104, USA. joshua.cooper@uphs.upenn.edu
Abstract:
A biventricular implantable cardioverter-defibrillator failed to detect monomorphic ventricular tachycardia (VT) with a cycle length that was within the programmed VT detection zone. VT beats were undetected because they fell within the atrial-paced ventricular blanking period while a rate smoothing pacing algorithm was turned on. This problem can occur with slow VT, a short AV delay, and a low upper tracking rate without any device interaction warning. This intradevice interaction can lead to untreated episodes of VT and result in recurrent episodes of heart failure and even death in patients with reduced left ventricular function and chronic heart failure.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias