Related Experiment Video
Updated: Aug 18, 2026

Three-Dimensional Modeling of the Left Atrium and Pulmonary Veins with a Precise Intracardiac Echocardiography Approach
Published on: June 30, 2023
Feasibility and validation of registration of three-dimensional left atrial models derived from computed tomography
Jasbir Sra1, David Krum, John Hare
1Electrophysiology Laboratories of Aurora Sinai and St. Luke's Medical Centers, University of Wisconsin Medical School-Milwaukee Clinical Campus, USA. bdanek@htcare.com
Objectives:
The purpose of this study was to determine the feasibility and assess the validity of registering three-dimensional (3D) models from computed tomographic (CT) images using a cardiac mapping system.
Background:
Registration of 3D anatomic models with an interventional system could help identify and navigate mapping and ablation catheters over a complex structure such as the left atrium (LA).
Methods:
ECG-gated, contrast-enhanced cardiac CT imaging was performed in 14 patients with atrial fibrillation. Segmentation was used to create 3D models of the LA. The 3D models were registered with the mapping system using a series of fiducial points. Registration was accomplished retrospectively in the first 10 patients, and catheter navigation was visualized from recorded data. In the final four patients, registration was accomplished in real time during electrophysiologic study. The mapping catheter position, as it was navigated inside the LA, was applied to the registered model in real time. For the validation study, temporary pacing leads were implanted in the LA of 10 dogs. Following this, CT scanning, segmentation, LA model importation, and registration was described previously. After registration, a mapping catheter was positioned at the site of each buried lead according to the registered model with no fluoroscopic guidance. A radiofrequency lesion was created at this location, and the dog was sacrificed, the heart removed and stained, and the distance between the buried lead and the lesion measured.
Results:
During the feasibility study, the location of the catheter in the registered model correlated with fluoroscopy, angiography, and intracardiac electrograms. LA endocardial potentials during sinus rhythm and any premature atrial contractions also were successfully delineated over the registered models. In the validation study, the mean target registration error was 2.0 +/- 3.6 mm.
Conclusions:
Registration of CT-derived 3D models of the LA using a cardiac mapping system is feasible and accurate.