Related Experiment Video
Updated: Aug 10, 2026

Measuring Nitrite and Nitrate, Metabolites in the Nitric Oxide Pathway, in Biological Materials using the Chemiluminescence Method
Published on: December 25, 2016
[Determination of nitrite in water samples by sequential injection analysis with chemiluminescence detection on chip
1Research Center for Analysis Sciences, Northeastern University, Shenyang 110004, China.
Abstract:
A sequential injection chemiluminescence's system using a chip flow cell was described for the determination of trace amount of nitrite in water samples. Nitride reacted with hydrogen peroxide to form peroxynitrous acid in sulfuric acid medium, which was an unstable in acid medium and subsequently was quenched into peroxynitrite in basic solution. During the composition of peroxynitrite the chemiluminescence intensity was enhanced with the presence of uranin and ethyldimethylcetylammonium bromide. The reaction coil (flow cell) and holding coil was combined for tracking the emission signal. The experimental parameters including physical and chemical parameter were optimized. The interference of cations in water samples was eliminated by passing previously the sample solution through a cation-exchange column. The linear range of the calibration graph was obtained from 1 x 10(-6) to 1 x 10(-4) mol x L(-1) with a correlation coefficient of 0.9998. The detection limit at 95% confidence was 6.8 x 10(-7) mol x L(-1). The relative standard deviation at 1 x 10(-5) mol x L(-1) nitrite level was 2.7%, and the recovery of 90%-99% and the sampling frequency of 80 h(-1) were obtained.