Marco Das1, Andreas Horst Mahnken, Georg Mühlenbruch
1Department of Diagnostic Radiology, University Hospital, University of Technology (RWTH), Pauwelsstrasse 30, D-52074 Aachen, Germany. das@rad.rwth-aachen.de
Objective:
The purpose of our study was to develop a simple protocol for reduction of radiation exposure without loss of diagnostic information in chest 16-MDCT.
Materials And Methods:
Two hundred and four patients underwent MDCT of the thorax (Somatom Sensation 16, Siemens). Group 1 was scanned using a standard protocol with 100 mAs(effective) (mAs(eff)). Group 2 was scanned using a dose modulation template (CareDose). Group 3 was scanned with mAs(eff) = body weight (kg). Group 4 was scanned with a combination of weight-adapted mAs(eff) and dose modulation. All other parameters were kept constant. Signal-to-noise ratio was assessed as an objective measurement for image quality, and subjective image quality was rated by three experienced radiologists on a 4-point scale. Effective dose was calculated using dedicated software.
Results:
The mean noise measurement values were 8.31 H for the 100 mAs(eff) protocol for the regression between weight and signal-to-noise (p < 0.0001), 9.08 H for group 2 (p < 0.0001), 9.0 H for group 3 (p = 0.5051), and 9.98 H for group 4 (p = 0.0152). The median image quality was 1 (1 = highest quality) in all subgroups. The mean effective dose was 6.83 mSv, 5.92 mSv, 4.73 mSv, and 3.97 mSv, respectively. The least correlation between weight and image noise was achieved for the individually weight-adapted protocol and in the weight-adapted with CareDose combination.
Conclusion:
By tube current time product adaptation (kg = mAs(eff)) combined with an online tube current modulation template, a well-balanced examination without significant loss of information was achieved for this specific scanner. Thus, individually adapted protocols for chest 16-MDCT can be recommended.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...