Related Experiment Video
Updated: Aug 12, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Separation of water-soluble vitamins by reversed-phase high performance liquid chromatography with ultra-violet
Olivier Heudi1, Tamara Kilinç, Patric Fontannaz
1Quality and Safety Department, Nestle Research Center, Nestec Ltd., Vers-Chez-les-Blanc, 1000 Lausanne 26, Switzerland. olivier.heudi@rdls.nestle.com
Abstract:
Nine water-soluble vitamins: [thiamine (B1), ascorbic acid (C), nicotinamide (PP), pyridoxine (B6), calcium pantothenate (B5), folic acid (B9), cyanocobalamin (B12), riboflavin (B2) and biotin (B8)] were separated on a YMC-Pack Pro C18 column (250 mm x 4.6 mm, 5 microm particle size) in a single run with a gradient elution of mobile phase consisting of 0.025% trifluoroacetic acid pH 2.6 (solvent A) and acetonitrile (solvent B). The separation was achieved within 17 min with a flow rate of 0.8 ml min(-1) and the detection was performed at two wavelengths (210 and 275 nm). The calibration graphs plotted with six concentrations of each vitamin were linear with a regression coefficient R2 > 0.995. The method was applied for the quantification of vitamins B1, C, PP, B6, B5, B9 B2 and B8 in polyvitaminated premixes (premixes) used for the fortification of infant nutrition products. The sample preparation involves an aqueous extraction of vitamins and two different samples dilution were used prior the LC-analysis. The specificity of the method was demonstrated by the retention characteristics, UV spectra and by comparing the peak purity with the standard of each vitamin. The repeatability of the method was evaluated at different level of concentrations on 12 premixes and the coefficients of variation (CVr) were below 6.5%. The values of the intermediate precision (CV1) were below 9.6% (n = 6). The concentrations of vitamins found in premixes with our method were comparable to the declared values, since no bias was found between the two sets of results at 95% confidence. The simplicity of the procedure should make it highly desirable for quality control of premixes in the food industry.
More Related Videos
10:22Improved UPLC-UV Method for the Quantification of Vitamin C in Lettuce Varieties (Lactuca sativa L.) and Crop Wild Relatives (Lactuca spp.)
Published on: June 30, 2020
05:03Quantitative Analysis of Dietary Vitamin A Metabolites in Murine Ocular and Non-Ocular Tissues Using High-Performance Liquid Chromatography
Published on: December 27, 2024
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
Ion-Exchange Chromatography
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...