Dual linear ion trap/orthogonal acceleration time-of-flight mass spectrometer with improved precursor ion selectivity
Yuichiro Hashimoto1, Hideki Hasegawa, Izumi Waki
1Central Research Laboratory, Hitachi, Ltd., Kokubunji-shi, Tokyo, Japan. h-yuichi@crl.hitachi.co.jp
Abstract:
A new hybrid mass spectrometer based on dual linear ion traps (LITs) and an orthogonal acceleration time-of-flight mass spectrometer (oaTOF), that can achieve MS(n) analysis and high-mass-accuracy detection with high sensitivity, has been developed. Dual-LIT was necessary because, in a single LIT plus oaTOF combination, the LIT pressure favorable for high precursor selectivity in MS(n) analysis (less than 1 mTorr) was far different from an optimum damping pressure (50-100 mTorr) for efficient connection to the TOF mass spectrometer. A dual-LIT solved this problem of inconsistency of the optimum pressures by using the first LIT for MS(n) analysis and the second LIT for collisional damping. This dual-LIT/TOF instrument achieved high-sensitivity MS(n) analysis with high precursor-ion selectivity.
Related Concept Videos
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrum: Interpretation
Mass Analyzers: Overview
Mass Analyzers: Common Types
Tandem Mass Spectrometry