Method for analysis of polar volatile trace components in aqueous samples by gas chromatography
Johan Pettersson1, Johan Roeraade
1Royal Institute of Technology, Department of Analytical Chemistry, SE-100 44 Stockholm, Sweden.
Abstract:
A new method has been developed for direct analysis of volatile polar trace compounds in aqueous samples by gas chromatography. Water samples are injected onto a short packed precolumn containing anhydrous lithium chloride. A capillary column is coupled in series with the prefractionation column for final separation of the analytes. The enrichment principle of the salt precolumn is reverse to the principles employed in conventional methods such as SPE or SPME in which a sorbent or adsorbent is utilized to trap or concentrate the analytes. Such methods are not efficient for highly polar compounds. In the LiCl precolumn concept, the water matrix is strongly retained on the hygroscopic salt, whereas polar as well as nonpolar volatile organic compounds show very low retention and are eluted ahead of the water. After transfer of the analytes to the capillary column, the retained bulk water is removed by backflushing the precolumn at elevated temperature. For direct injections of 120 microL of aqueous samples, the combined time for injection and preseparation is only 3.5 min. With this procedure, direct repetitive automated analyses of highly volatile polar compounds such as methanol or tetrahydrofuran can be performed, and a limit of quantification in the low parts-per-billion region utilizing a flame ionization detector is demonstrated.
More Related Videos
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
05:29Profiling Volatile Compounds in Blackcurrant Fruit using Headspace Solid-Phase Microextraction Coupled to Gas Chromatography-Mass Spectrometry
Published on: June 9, 2021
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Volatilization
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Voltammetric Techniques: Linear-Scan (E vs Time)
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...