Cone-beam-CT guided radiation therapy: A model for on-line application
Mark Oldham1, Daniel Létourneau, Lindsay Watt
1Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, MI, USA.
Summary
On-line X-ray volumetric cone-beam CT imaging (XVI) improves radiation therapy accuracy by enabling precise patient set-up. This study demonstrates XVI
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Image-Guided Therapy
Background:
- Accurate patient positioning is critical for effective radiation therapy.
- Traditional set-up methods can be limited by anatomical variations and motion.
- On-line imaging offers a potential solution for real-time error correction.
Purpose of the Study:
- To present efficient and generalized processes for clinical application of on-line X-ray volumetric cone-beam CT imaging (XVI).
- To improve the accuracy of patient set-up in radiation therapy using XVI.
- To illustrate XVI image-guided therapy in contrasting sites: intra-cranial radiosurgery and prostate radiation therapy.
Main Methods:
- A two-step process for determining on-line set-up errors was developed.
- Step 1: Registration of XVI data to planning data by matching machine and planning isocenters.
- Step 2: Second registration to correct residual displacements; accuracy investigated in phantom studies.
Main Results:
- Sub-pixel set-up errors (down to 0.5mm) were accurately determined and implemented in radiosurgery.
- XVI provided quality skull detail for precise on-line skull-based alignment.
- A 'local XVI' technique showed soft-tissue detail for on-line prostate set-up.
Conclusions:
- XVI has significant potential to enhance radiation treatment accuracy.
- Current XVI image quality supports bony and soft-tissue set-up error determination and correction.
- Efficient procedures for utilizing on-line XVI data are crucial for image-guided therapy.