Related Experiment Video
Updated: Aug 18, 2026

Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
Computed tomography colonography: automated diameter and volume measurement of colonic polyps compared with a manual
David Burling1, Steve Halligan, Mary E Roddie
1Intestinal Imaging Centre, St. Mark's Hospital, Northwick Park, London, United Kingdom.
Objective:
To investigate inter- and intraobserver agreement of automated measurement of polyp diameter in vitro.
Methods:
Two phantoms ("QRM" and "Whiting") containing simulated polyps of known diameter and volume were scanned using 16-detector row computed tomography. Two observers estimated polyp diameter using 3 methods: software calipers ("manual"), freehand boundary identification ("semiautomatic"), and automated software segmentation ("fully automatic").
Results:
Intraobserver 95% limits of agreement for diameter were narrowest for the fully automatic method (QRM span: 0.39 mm, 0.48 mm; Whiting span: 0.24 mm, 0 mm). Manual estimates were approximately 10 times wider (QRM span: 3.57 mm, 3.21 mm; Whiting span: 3.2 mm, 2.02 mm). Volume estimates were narrowest for the fully automatic method (span: 24.2 mm, 24.1 mm vs. 97.9 mm, 102.9 mm for semiautomatic measurement). Interobserver agreement for diameter was narrowest for the fully automatic method (QRM span: 0.12 mm, Whiting span: 0.16 mm), with the manual method approximately 18 times wider (QRM span: 2.87 mm, Whiting span: 2.18 mm).
Conclusion:
Fully automated measurement of polyp diameter and volume is technically feasible and results in superior inter- and intraobserver agreement.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Endoscopic Procedures II: Colonoscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...