Dosimetry of digital panoramic imaging. Part I: Patient exposure
F Gijbels1, R Jacobs, R Bogaerts
1Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Katholieke Universiteit Leuven, Kapucijnenvoer 7, 3000 Leuven, Belgium.
Objectives:
To measure patient radiation dose during panoramic exposure with various panoramic units for digital panoramic imaging.
Methods:
An anthropomorphic phantom was filled with thermoluminescent dosemeters (TLD 100) and exposed with five different digital panoramic units during ten consecutive exposures. Four machines were equipped with a direct digital CCD (charge coupled device) system, whereas one of the units used storage phosphor plates (indirect digital technique). The exposure settings recommended by the different manufacturers for the particular image and patient size were used: tube potential settings ranged between 64 kV and 74 kV, exposure times between 8.2 s and 19.0 s, at fuse current values between 4 mA and 7 mA. The effective radiation dose was calculated with inclusion of the salivary glands.
Results:
Effective radiation doses ranged between 4.7 microSv and 14.9 microSv for one exposure. Salivary glands absorbed the most radiation for all panoramic units. When indirect and direct digital panoramic systems were compared, the effective dose of the indirect digital unit (8.1 microSv) could be found within the range of the effective doses for the direct digital units (4.7-14.9 microSv).
Conclusions:
A rather wide range of patient radiation doses can be found for digital panoramic units. There is a tendency for lower effective doses for digital compared with analogue panoramic units, reported in previous studies.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...