Related Experiment Video
Updated: Aug 18, 2026

Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Pitfall in the high-throughput quantification of whole blood cyclosporin A using liquid chromatography-tandem mass
1Institute of Clinical Chemistry, Hospital of the University of Munich, Germany. michael.vogeser@med.uni-muenchen.de
Abstract:
In a growing number of laboratories the technique of liquid chromatography-tandem mass spectrometry is used for the quantification of cyclosporin A in whole blood, employing cyclosporin D as the internal standard. Cyclosporin A is extensively metabolized in vivo; in liquid chromatography-tandem mass spectrometry respective metabolites can give rise to both parent and product ions that are isobaric with ions commonly used for the detection of cyclosporin A and cyclosporin D, respectively. In this article it is demonstrated that limited chromatography with co-elution of such metabolites together with cyclosporin A and cyclosporin D can lead to incorrect results.