Objectives:
The objectives of this study were to assess practicability, outcome, and possible advantages of magnetic resonance imaging (MRI)-guided localizations with modified embolization coils for exclusively MRI-detected suspicious breast lesions.
Materials And Methods:
We performed 131 MRI-guided preoperative localizations in 105 patients with exclusively MRI-detected breast lesions. Contrast-enhanced, T1-weighted, 3-dimensional gradient echo sequences were used for imaging. Breast fixation and needle guidance were achieved by a perforated plate. Lesion localization was performed with a modified embolization coil. The distance between coil and lesion was measured. Results of specimen radiography and histopathology were analyzed.
Results:
Thirty-four of 131 lesions (26%) were malignant. Lesion localization and excision was successful in all patients. The mean distance between the coil and the lesion was 0.5 mm (+/-1.7 mm standard deviation).
Conclusions:
Exact MRI-guided preoperative localization of exclusively MRI-detected breast lesions by using a modified embolization coil is feasible with excellent precision. Advantages over wire localization procedures are the possibility of specimen x-ray and the independence from time and place of operation.
More Related Videos
05:52Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
Published on: November 4, 2021
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013