Absolute CBF and CBV measurements by MRI bolus tracking before and after acetazolamide challenge: repeatabilily and

Cécile B Grandin1, Anne Bol, Anne M Smith

  • 1Department of Radiology, St Luc University Hospital, Université Catholique de Louvain, Brussels, Belgium. grandin@rdgn.ucl.ac.be

Neuroimage
|May 24, 2005
PubMed

Insights

Bolus tracking MRI offers moderate accuracy for cerebral blood flow (CBF) and volume (CBV) in healthy humans. Adjusting MRI values with a scale factor may improve agreement with PET imaging for better clinical use.

Area of Science:

  • Neuroimaging
  • Medical Physics
  • Radiology

Background:

  • Bolus tracking MRI is used for cerebral perfusion assessment.
  • Its accuracy in measuring absolute cerebral blood flow (CBF) and cerebral blood volume (CBV) is debated.
  • Validation against positron emission tomography (PET) is needed.

Purpose of the Study:

  • To validate echo planar gradient echo MRI for absolute CBF and CBV measurements.
  • To compare MRI-derived values with H(2)(15)O and (11)CO PET in healthy humans.
  • To assess MRI's utility in evaluating cerebrovascular reserve.

Main Methods:

  • Comparison of MRI and PET measurements of CBF and CBV in 12 healthy subjects.
  • Measurements were taken at rest and after acetazolamide or saline administration.
  • Repeatability and agreement between MRI and PET were analyzed.

Main Results:

  • Moderate repeatability for both MRI and PET, slightly lower for MRI.
  • Similar mean CBF values between MRI and PET, except in the cortex (higher with MRI).
  • CBV was consistently higher with MRI, potentially due to arterial input function underestimation.
  • MRI and PET showed significant CBF increase after acetazolamide, indicating potential for cerebrovascular reserve assessment.
  • Good individual subject correlation (R(2) 0.70-0.84) but lower group correlation (R(2) 0.40-0.65) with large agreement limits.

Conclusions:

  • MRI may provide physiologic CBF values but systematically overestimates CBV in healthy humans.
  • An individual scale factor applied to MRI measurements could improve agreement with PET.
  • MRI shows potential for assessing cerebrovascular reserve, but calibration is necessary.