Related Experiment Video
Updated: Aug 17, 2026

Large-scale Top-down Proteomics Using Capillary Zone Electrophoresis Tandem Mass Spectrometry
Published on: October 24, 2018
Over 1000-fold protein preconcentration for microliter-volume samples at a pH junction using capillary
Chandra A Nesbitt1, Josephine T M Lo, Ken K C Yeung
1Department of Chemistry, The University of Western Ontario, London, Ont., N6A 5B7 Canada.
Abstract:
An effective protein preconcentration technique specifically designed for microliter-volume samples is presented. The preconcentration is based on the capturing of protein in its isoelectric point (pI) within an applied electric field, using a pH junction created by a discontinuous buffer system. The buffers were chosen to selectively preconcentrate proteins of neutral pI, myoglobin in this case, while removing other proteins with acidic or basic pIs. For the suppression of electro-osmotic flow (EOF) and protein adsorption, the capillary inner wall was modified with a zwitterionic phospholipid bilayer coating. A preconcentration factor of up to 1700 was obtained for a 1 microg/mL solution of myoglobin. The preconcentration was completed in approximately 20 min. Several sample introduction conditions were presented to accommodate sample volume from one to a few hundreds of microliters. The final volume of the preconcentrated sample band was estimated to be approximately 5 nL.
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...