Validation of traction force indicator in ventouse devices
B J Whitlow1, O Tamizian, J Ashworth
1Colchester Hospital, Turner Road, Colchester CO4 5JL, UK. barry.whitlow@essexrivers.nhs.uk
Objective:
To evaluate whether the traction force indicator in a ventouse device gave an accurate estimation of the force applied.
Methods:
The study was carried out at the Derby City Hospital. An estimate of measurement errors was made by analyzing the standard deviation of the residuals for 30 Kiwi OmniCup devices (Clinical Innovations, Abingdon, Oxfordshire, England) and a standard industrial spring balance.
Results:
The measurement errors for the traction forces were very small and acceptable for the OmniCup. The overall estimate of measurement errors was 0.45 kg (1.35 lb).
Conclusion:
The overall measurement error for the traction force indicator of the OmniCup was found to be very small. Injuries associated the ventouse, such as cephalohematomas and subgaleal hemorrhages, may be minimized if the recommended limits for a safe traction force are not exceeded.
Related Concept Videos
Static and Kinetic Frictional Force
However, if two systems are in contact and are stationary relative to one...
Net Torque Calculations
Magnetic Force
The magnetic force acting on a moving charge...
Centrifugal Force
Power Expended by a Constant Force
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...