X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Varied tube potential with constant effective dose at lumbar spine radiography using a flat-panel digital detector
1Department of Radiology, Orebro University Hospital, SE-701 85 Orebro, Sweden. hakan.geijer@orebroll.se
Abstract:
The purpose of the study was to evaluate the image quality at different tube potential (kV) settings using anteroposterior lumbar spine radiography as a model. An Alderson phantom was used with a flat-panel detector. The tube potential varied between 48 and 125 kV while the tube charge (mAs) was adjusted to keep an effective dose of 0.11 mSv. Image quality was assessed with a visual grading analysis and with a CDRAD contrast-detail phantom together with a computer program. The VGA showed inferior image quality for the higher kV settings, > or =96 kVwith similar results for the contrast-detail phantom. When keeping the effective dose fixed, it seems beneficial to reduce kV to get the best image quality despite the fact that the mAs is not as high as with automatic exposure. However, this cannot be done with automatic exposure, which is set for a constant detector dose.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET