A survey of patient dose and clinical factors in a full-field digital mammography system
Related Experiment Video
Updated: Aug 5, 2026

Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
P Morán1, M Chevalier, J I Ten
1Medical Physics, Department of Radiology, Facultad de Medicina, Universidad Complutense de Madrid, 28040 Madrid, Spain.
Abstract:
In this work, we analyse the dose to 5034 patients (20 137 images) who underwent mammographic examinations with a full-field digital mammography (FFDM) system over a 2-y period. The information relevant to this study has been extracted from the image Digital Imaging and Communications in Medicine (DICOM) headers. Entrance surface air-kerma (ESAK) without backscatter and average glandular dose (AGD) were estimated following the methodology proposed in the European Protocol on Dosimetry in Mammography. Mean values for patient age and compressed breast thickness were 56 +/- 11 y and 52 +/- 13 mm, respectively. The mean ESAK value was 8.1 mGy and the mean AGD was 1.9 mGy. In addition, the dose values from both FFDM and screening-film mammographic (SFM) examinations were compared. The third quartile (TQ) of the ESAK values delivered by the FFDM system was 33% lower and 32% higher than the TQ for SFM with slow and fast screen/film receptors. Differences between dose values for cranio-caudal (CC) and medio-lateral oblique (MLO) images (about 27% for SFM) decreased to 11% for FFDM.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...