Related Experiment Video
Updated: Aug 17, 2026

Bergmeyer Glucose Quantification for Microbiological Samples
Published on: January 17, 2025
Evaluation of a glucose meter for determining the glycemic responses of foods
Amogh Velangi1, Glen Fernandes, Thomas M S Wolever
1Department of Nutritional Sciences, University of Toronto, Toronto, ON, Canada M5S 3E2.
Background:
Glucose meters are convenient for measuring postprandial glycemic responses. However, their performance for this purpose has not been evaluated.
Methods:
Glucose responses of 7 potato meals were measured using the One Touch Ultra(R) (OTU) glucose meter and a reference method (Yellow Springs Instruments Glucose Analyzer, YSI) and the incremental areas under the curves (AUC) and glycemic index (GI) values compared.
Results:
Mean AUC(OTU) was greater than AUC(YSI) (192+/-8 vs. 175+/-7 mmolxmin/l, p=0.001), but GI(OTU) tended to be less than GI(YSI) (69+/-3 vs. 74+/-3, p=0.052). Bland-Altman difference plots showed wide 95% limits of agreement for AUC (-84 to 119 mmolxmin/l) and GI (-21 to 26) values of individual subjects, and for the mean GI values of the 7 foods (-11 to 21). Total and error variance of AUC and GI values were greater for OTU than YSI, and food means differed significantly for YSI (p<0.01) but not OTU (p=0.11).
Conclusion:
AUC and GI values determined by OTU are more variable and do not agree well with those obtained by YSI. Thus, the OTU is not recommended for determining AUC or GI in normal subjects. This conclusion does not necessarily apply to other glucose meters whose performance should be evaluated.
Related Concept Videos
Hyperglycemia
Hypoglycemia and Glucagon
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Overview of Carbohydrate Metabolism
Glucose transport into cells is facilitated by a family of transport proteins called GLUT (Glucose Transporters). GLUT4 is the primary glucose transporter for insulin-stimulated glucose...