High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Proficiency testing (external quality assessment) of drug detection in oral fluid
1Altrix Healthcare plc, Birchwood Science Park, Warrington WA3 7BP, UK. j.clarke@altrix.com
Abstract:
Eighteen external quality assessment (proficiency testing) samples were prepared from client specimens collected with the Intercept oral fluid collection device and by spiking drug-free oral fluid. Samples were circulated in pairs at quarterly intervals to 13 UK and USA based laboratories for analysis by a panel of OraSure micro-plate Intercept enzyme immunoassay kits and hyphenated mass spectrophotometric techniques. During the survey, there was a single case of non-specificity in a false report for methadone. The major errors were of lack of sensitivity relative to the concentration thresholds specified for the immunoassays. The sensitivity for overall "present"/"not found" reports calculated as true positives/(true positives+false negatives) were for the amphetamine specific assay 50%, methyl-amphetamines 93%, barbiturates 64%, cannabinoids 73%, cocaine and metabolites 100%, benzodiazepines 69%, methadone 95%, opiates 79% (opiates excluding oxycodone 93%), phencyclidine 93% and human gamma-globulin 97%. A small number of the sensitivity errors were attributable to errors in chromatographic confirmation techniques.