Electrode interaction in pediatric cochlear implant subjects.

Marc D Eisen1, Kevin H Franck

  • 1Center for Childhood Communication, Children's Hospital of Philadelphia, Philadelphia, PA, USA. eisenm@mail.med.upenn.edu

Journal of the Association for Research in Otolaryngology : JARO
|June 14, 2005
PubMed
Summary

Electrode interaction in pediatric cochlear implants increases with stimulus intensity and varies by electrode location. Understanding this interaction is key for improving speech perception with cochlear implants.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Voluntary Field Recall of Advanced Bionics HiRes Cochlear Implants: A Single-Institution Experience.

Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology·2022
Same author

Fractionated Proton Radiation Therapy and Hearing Preservation for Vestibular Schwannoma: Preliminary Analysis of a Prospective Phase 2 Clinical Trial.

Neurosurgery·2022
Same author

Implementation of Mobile Audiometry During the COVID-19 Pandemic.

Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery·2021
Same author

Patient Report of Hearing in Neurofibromatosis Type 2: Recommendations for Clinical Trials.

Neurology·2021
Same author

Predicting Depression From Hearing Loss Using Machine Learning.

Ear and hearing·2021
Same author

Regulation of Over-the-Counter Hearing Aids - Deafening Silence from the FDA.

The New England journal of medicine·2020

Area of Science:

  • Audiology
  • Biomedical Engineering
  • Neuroscience

Background:

  • Cochlear implants use differential electrode stimulation for speech processing.
  • Electrode interaction, or nerve fiber overlap, reduces spectral resolution in cochlear implant users.
  • This study investigates electrode interaction in pediatric cochlear implant subjects.

Purpose of the Study:

  • To define electrode interaction in pediatric cochlear implant subjects.
  • To assess the influence of stimulus intensity and electrode location on electrode interaction.
  • To inform future cochlear implant array design and fitting strategies.

Main Methods:

  • Electrically evoked compound action potentials (ECAPs) were measured in 27 pediatric subjects.
  • A forward masking protocol was employed to estimate electrode interaction.

Related Experiment Videos

  • Interaction was quantified by comparing probe responses with and without a masker electrode.
  • Main Results:

    • Electrode interaction significantly increased with higher stimulus intensity (p < 0.05).
    • Interaction was lower at basal electrode locations compared to middle and apical locations (p < 0.05).
    • A significant correlation was found between electrode distance from the base and interaction levels (p < 0.001).

    Conclusions:

    • Electrode interaction in pediatric cochlear implant users is dependent on both stimulus intensity and electrode position.
    • Findings suggest implications for optimizing cochlear implant array design and current fitting protocols.
    • The precise impact of electrode interaction on overall implant performance requires further investigation.