Survivorship of the Charnley Elite Plus cemented femoral stem
M Rowsell1, J Der Tavitian, S Birtwistle
1Glenfield Hospital, Leicester, England, UK. mark.rowsell@doctors.org.uk
Insights
The Charnley Elite Plus femoral stem showed a 1.9% revision rate for aseptic loosening and a 5.8% radiographic failure rate within 4.5 years. Early failures raise concerns about the long-term performance of this hip arthroplasty implant.
Area of Science:
- Orthopedic Surgery
- Biomaterials Engineering
- Medical Device Performance
Background:
- Total hip arthroplasty (THA) is a common procedure for hip joint degeneration.
- Femoral stems are critical components influencing THA success and longevity.
- The Charnley Elite Plus femoral stem is a widely used implant in primary hip arthroplasty.
Purpose of the Study:
- To evaluate the early clinical and radiographic outcomes of the Charnley Elite Plus femoral stem.
- To assess the survivorship and failure modes of this prosthesis in a multi-surgeon setting.
- To identify potential concerns regarding the long-term performance based on early results.
Main Methods:
- Prospective study of 268 primary hip arthroplasty procedures using the Charnley Elite Plus stem.
- Minimum follow-up of 3 years, with postal and radiographic review at a mean of 4.5 years.
- Analysis of revision procedures for aseptic loosening and radiographic evidence of stem failure.
Main Results:
- Five revision procedures (1.9%) were performed due to aseptic loosening.
- Twelve femoral stems (5.8%) showed gross radiological failure.
- Survivorship was 98.1% when only considering revision for aseptic loosening, but decreased to 93.1% when including radiographic failures.
Conclusions:
- The Charnley Elite Plus femoral stem demonstrated a notable rate of early aseptic loosening and radiographic failure.
- The observed failure rate at a mean of 4.5 years raises significant concerns regarding the long-term survivorship of this implant.
- Further long-term monitoring is warranted to fully understand the performance profile of the Charnley Elite Plus femoral stem.
Abstract:
We report the results of the Charnley Elite Plus femoral stem in multiple surgeons' hands at a minimum of 3 years after implantation. Over a 4-year period, 244 patients underwent 268 primary hip arthroplasty procedures using this particular stem. Patients underwent postal and radiographic review at a mean of 4.5 (3.0--6.8) years following their arthroplasty. There were five revision procedures for aseptic loosening (5/268; 1.9%). Radiological assessment revealed gross radiological failure in a further 12 femoral stems (12/208; 5.8%). In the best case scenario, using revision for aseptic loosening as the endpoint, the survivorship for this period was 98.1%. If radiographic failures are incorporated into this endpoint, survivorship is 93.1%. This failure rate at an early stage raises concerns as to the long-term survivorship of this prosthesis.