In vitro properties of PLLA screws and novel bioabsorbable implant with elastic nucleus to replace intervertebral
V Ellä1, M Kellomäki, P Törmälä
1Institute of Biomaterials, Tampere University of Technology, P.O. Box 589, FIN-33101, Tampere, Finland, ville.ella@tut.fi.
Abstract:
The suitability of two different implant types for the replacement of the intervertebral disc was studied in vitro. Self-reinforced poly-L-lactide (SR-PLLA) screws Ø 4.5 mm were studied 24 weeks in vitro and cylindrical implants with elastic nucleus made of poly(L/D)lactide 96/4, poly(L/DL)lactide 70/30, Bioactive Glass n:o 13-93 and Polyactive 1000PEOT70PBT30 were studied 15 weeks in vitro. The cylindrical implant mimics the size and shape of the intervertebral disc. During the in vitro, there were no changes in compression properties with either implant types. The screws had sufficient modulus for intervertebral ossification in the canine model and the cylindrical implant showed also sufficient mechanical properties. These results suggest that both implant types could be used in clinical testing.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
07:06Surgical Technique for the Implantation of a Biomimetic Artificial Intervertebral Disc in a Goat Animal Model
Published on: October 10, 2025
Related Concept Videos
Herniated Intervertebral Disc l: Introduction
Degenerative Disc Disease I: Introduction
Degenerative Disc Disease ll: Pathophysiology