Abstract:
We developed and validated an HPLC method with intramolecular excimer-forming fluorescence derivatization to determine methylmalonic acid, a unique biochemical marker for methylmalonic aciduria. Methylmalonic acid in urine and an internal standard were derivatized with pyrenebutyric hydrazide and separated on a C8 column. The derivatives were detected by monitoring the fluorescence at 475 nm (excitation wavelength 345 nm). At a signal-to-noise ratio of 3, the detection limit was 0.33 pmol on the column and the calibration curve was linear up to 1 mmol[sol ]L in urine. In a retrospective study on a relatively large number of known methylmalonic aciduria cases (n = 48), the method enabled us to differentiate methylmalonic aciduria cases from healthy controls (n = 52), regardless of age of patients at sampling or years of specimen storage. No interference was observed from isomeric or other dicarboxylic acids, or other urine constituents. As described, the method can be used retrospectively or prospectively for the diagnosis of methylmalonic aciduria and can be easily adopted by laboratories with no access to gas chromatography-mass spectrometry.
More Related Videos
11:53An Alternative Culture Method to Maintain Genomic Hypomethylation of Mouse Embryonic Stem Cells Using MEK Inhibitor PD0325901 and Vitamin C
Published on: June 1, 2018
09:40A Mass Spectrometry-Based Proteomics Approach for Global and High-Confidence Protein R-Methylation Analysis
Published on: April 28, 2022