Related Experiment Video
Updated: Aug 8, 2026

Determination of the Relative Potency of an Anti-TNF Monoclonal Antibody (mAb) by Neutralizing TNF Using an In Vitro Bioanalytical Method
Published on: September 16, 2017
A one-step, competitive electrochemiluminescence-based immunoassay method for the quantification of a fully human
D Horninger1, E Eirikis, C Pendley
1Department of Clinical Pharmacology and Experimental Medicine, Centocor, Inc., 145 King of Prussia Road, Radnor, PA 19087, USA.
Abstract:
A quantitative, one-step, competitive electrochemiluminescence (ECL)-based immunoassay for the determination of a fully human, anti-TNFalpha monoclonal antibody in human serum has been developed. A biotinylated, mouse anti-variable region-specific antibody and a ruthenium-labeled anti-TNFalpha antibody were the only specific reagents needed to develop the assay. A single incubation step of 2 h followed by ECL detection was used. The assay was capable of measuring the analyte in neat serum over approximately a 1600-fold range with higher concentrations measured following a single dilution. Assay accuracy, precision, and reproducibility were suitable to support pharmacokinetic studies of the analyte. This competitive assay format offers an alternative approach to the development of immunoassays for the measurement of macromolecules in complex matrices to support preclinical and clinical studies.