Related Experiment Video
Updated: Jan 12, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Evolution of the longitudinal magnetization for pulse sequences using a fast spin-echo readout: application to
Simon J P Meara1, Gareth J Barker
1Centre for Neuroimaging Sciences, Department of Neurology, Institute of Psychiatry, London, United Kingdom. s.meara@iop.kcl.ac.uk
Abstract:
The fast spin-echo (FSE) sequence is frequently used as a fast data-readout technique in conjunction with other pulse sequence elements, such as in fluid-attenuated inversion-recovery (FLAIR) and double inversion-recovery (DIR) sequences. In order to implement those pulse sequences, an understanding is required of how the longitudinal magnetization evolves during the FSE part of the sequence. This evolution has been addressed to a certain extent by previous publications, but the DIR literature in particular appears to be replete with approximations to the exact expression for the longitudinal magnetization, and several papers contain errors. Equations are therefore presented here for the evolution of the longitudinal magnetization for a FSE readout. These are then applied to calculate the magnetization available immediately prior to the 90 degrees imaging pulse for the FLAIR-FSE and DIR-FSE pulse sequences.